H serise fully-automatic video measuring machine

Short Description:

H series automatic video measuring machine adopts HIWIN P-level linear guide, TBI grinding screw, Panasonic servo motor, high-precision metal grating ruler and other precision accessories. With an accuracy of up to 2μm, it is the measurement device of choice for high-end manufacturing. It can measure 3D dimensions with an optional Omron laser and Renishaw probe.We customize the height of the Z axis of the machine according to your requirements.


  • Measurement Range: 400*300*200mm
  • Measurement Accuracy: 2.5+L/200
  • Optical Magnification: 0.7-4.5X
  • Image Magnification: 30-200X
  • Product Detail

    Product Tags

    Main Technical Parameters and Characteristics of the Machine

    Model

    HD-322H

    HD-432H

    HD-542H

    Overall dimensions(mm)

    550×970×1680mm

    700×1130×1680mm

    860×1230×1680mm

    X/Y/Z axis Range(mm)

    300×200×200

    400×300×200

    500×400×200

    Error of indication(um)

    E1(x/y)=(2.5+L/100)

    Workbench load(kg)

    25kg

    Instrument weight(kg)

    240kg

    280kg

    360kg

    Optical System

    CCD

    1/2”CCD industrial color camera

    Objective lens

    Automatic zoom lens

    Magnification

    Opital Magnification:0.7X-4.5X;Image Magnification:24X-190X

    Working distance

    92mm

    Object field of view

    11.1 ~1.7mm

    Grating resolution

    0.0005mm

    Transmission system

    HIWIN P-level linear guide, TBI grinding screw

    Motion Control System

    Panasonic CNC Servo Motion Control System

    Speed

    XY axismm/s)

    200

    axismm/s)

    50

    Light source system

    The surface light adopts 5-ring and 8-zone LED cold light source, and each section is controlled independently; the contour light is an LED transmission parallel light source, and 256-level brightness can be adjusted

    Measurement software

    INSPECT 3D software

    Working Environment of Instrument

    H serise

    Temperature and humidity
    Temperature: 20-25℃, optimal temperature: 22℃; relative humidity: 50%-60%, optimal relative humidity: 55%; Maximum temperature change rate in the machine room: 10℃/h; It is recommended to use a humidifier in dry area, and use a dehumidifier in humid area.

    Heat calculation in the workshop
    ·Keep the machine system in the workshop operating in the optimum temperature and humidity, and the total indoor heat dissipation must be calculated, including the total heat dissipation of indoor equipment and instruments (lights and general lighting can be ignored)
    ·Heat dissipation of human body: 600BTY/h/person
    ·Heat dissipation of workshop: 5/m2
    ·Instrument placement space (L*W*H): 3M ╳ 2M ╳ 2.5M

    Dust content of air
    The machine room shall be kept clean, and the impurities greater than 0.5MLXPOV in the air shall not exceed 45000 per cubic foot. If there is too much dust in the air, it is easy to cause resource read and write errors and damage to the disk or read-write heads in the disk drive.

    Vibration degree of machine room
    The vibration degree of machine room shall not exceed 0.5T. Machines that vibrate in the machine room shall not be placed together, because the vibration will loosen the mechanical parts, joints and contact parts of the host panel, resulting in abnormal operation of the machine.

    Power Supply

    AC220V/50HZ

    AC110V/60HZ

    FAQ

    What is your company's QC standard?

    QC mechanical accuracy: XY platform indication value 0.004mm, XY verticality 0.01mm, XZ verticality 0.02mm, lens verticality 0.01mm, concentricity of magnification <0.003mm.

    How long is the service life of your products?

    Our equipment has an average lifespan of 8-10 years.

    Which groups and markets are your products suitable for?

    Our products are suitable for dimensional measurement in electronics, precision hardware, molds, plastics, new energy, medical equipment, automation equipment and other industries.